ProductsLocation:Home >Products

  • Advantages
  • Indicators
  • Contrast
  • Accessories
  • Instructions

Feature

High-voltage insulation resistance tester, with 5 range: 500V, 1000V, 2500V, 5000V, 10000V, 15000V20000V,test up to 20TΩ.

1.Strict accordance with the safety standards for design

2. Insulation resistance range 20TΩ @ 20 kv

3.Short circuit current can be adjusted, maximum 5mA (Can be ordered as high as 10mA, short circuit current products).

4. Automatic display polarization index (PI), induced electric absorption test values than (DAR), and can test the leakage current and capacitance.

5. Anti-jamming performance is superior, the interference current reaches 2 mA, still ensure accuracy of test instruments.

6. 2 methods for power: use of lithium ion batteries, battery life can be up to 4 hours (10000@ 100MΩ).

7. Insulation resistance simulation column.

8. Digital filtering functions, external influence deviation display value causing mitigation point using filter function.

9. Perfect protection function, and the spare fuse prompt functions.

● Perfect protection function, and the spare fuse prompt functions.


Products

HZJY-20KV insulation resistance tester

Advantages

Feature

High-voltage insulation resistance tester, with 5 range: 500V, 1000V, 2500V, 5000V, 10000V, 15000V20000V,test up to 20TΩ.

1.Strict accordance with the safety standards for design

2. Insulation resistance range 20TΩ @ 20 kv

3.Short circuit current can be adjusted, maximum 5mA (Can be ordered as high as 10mA, short circuit current products).

4. Automatic display polarization index (PI), induced electric absorption test values than (DAR), and can test the leakage current and capacitance.

5. Anti-jamming performance is superior, the interference current reaches 2 mA, still ensure accuracy of test instruments.

6. 2 methods for power: use of lithium ion batteries, battery life can be up to 4 hours (10000@ 100MΩ).

7. Insulation resistance simulation column.

8. Digital filtering functions, external influence deviation display value causing mitigation point using filter function.

9. Perfect protection function, and the spare fuse prompt functions.

● Perfect protection function, and the spare fuse prompt functions.


Indicators

Specification Parameter

1. Safety Standard

IEC 61010-1    CAT.IV  600V

IEC 61326      EMC specification: test, control and electrical equipment for testing

IEC60529 IP64 (Outer closed state)

2. AC Power:220V±10%,50/60 HZ ,20 VA

3. battery:16.8V Lithium ion rechargeable battery 

4. Battery Life:10000V@100M, 4 hour 

5. size(L H):37cm x 27cm x 17cm

6. weight:3.6kg

7.Test voltage accuracy: the nominal value  100% to 110%

8. Voltage measurement accuracy:5%+3V

9. Current test range:10mA

10. current measure precision:5%+0.2nA

11.short-circuit current: 5mA

12. Capacitance test range:20uF

13. Capacitance testing accuracy:15%+0.03uF

14.Insulation resistance testing range and accuracy (T:23±5º,RH :45 – 75%)

     range

 

accuracy

 

500V

1000V

2500V

5000V

10000V

Unspecified

<100k

<100k

<100k

<100k

<100k

5%

100k-10G

100k-20G

100k-100G

100k-100G

100k-200G

20%

10G -100G

20G-200G

100G-1T

100G-1T

200G-2T

Unspecified

> 100G

> 200G

> 1T

> 1T

> 2T


Contrast

Accessories

1. The high-pressure silica gel test line (red, green and black one)  1

2. AC 220V power line            1

3. Operating manual                     1

4. Certificate of soundness                      1

 


Instructions

All rights reserved © ICP15020314-1 Huazheng Electric Manufacturing (Baoding) Co., Ltd.
No.3088 Lekai Street, Baoding Tel:17731277786 Fax:0312-6775656