Primary Technical Specifications
1.The capacitance range of the test object 6pF--250μF
2.Measuring sensitivity and Max. current(see table 1)
No. Tuning capacitance range Sensitivity (pC)
(Un-balance circuit) RMS Of max. current
(Un-balance circuit) (Balance circuit)
1 0~25~100pF 0.02 30mA 0.25A
2 25~100~400pF 0.04 50mA 0.5A
3 100~400~1500pF 0.06 120mA 1A
4 400~1500~6000pF 0.1 0.25A 2A
5 1500~6000~25000pF 0.2 0.5A 4A
6 0.006~0.025~0.1μF 0.3 1A 8A
7 0.025~0.1~0.4μF 0.5 2A 15A
8 0.1~0.4~1.5μF 1 4A 30A
9 0.4~1.5~6.0μF 1.5 8A 60A
10 1.5~6.0~25μF 2.5 15A 120A
11 6.0~25~60μF 5 25A 200A
12 25~60~250μF 10 50A 300A
7R Resistor 0.5 2A 15A
Table 1
3. Band width:
(1) On low frequency 3dB, fL: 10kHz, 20kHz, 40 KHz optional
(2) On high frequency 3dB, fH: 80kHz, 200kHz, 300 KHz optional
4、Gain adjusting:
6 grades of rough gain. Gain between 2 near grades is 20±1dB, fine adjusting range >20dB
5、Time window:
① Window width: Adjustable within 15º~150º
② Window position: Rotatable within 0º~170º
③ Two windows can be set on/off respectively or simultaneously
6、Partial Discharge Meter:
Digital meter with 3.5 bits LED digital display,;
Partial discharge magnitude deviation within 0-100: <±5%
7、Ellipse time base:
① Frequency: 60 Hz, 120 Hz, 180 Hz, 240 Hz, 300Hz
② Ellipse rotation: 30º per grade, total 150º
③ Display mode: ellipse——line——sine
8、Test high voltage meter:
Range: 100kV (extensible)
② Display: 3½ digital voltage meter,precision better than ±5%(for full graduation)
9、It has zero-mark, which is useful to distinguish the partial discharge.
10、Size :460*420*260(Width*Length*Height)mm。
11、Weight: About 16kG.