CS2671AX Withstanding Voltage Tester
CS2671AX Withstanding Voltage Tester

I.what will you notice before testing

The output of the withstanding voltage tester can output 20kV, the highest voltage. Any false operation will result in the misadventure even death. So please read this chapter carefully with the view of safety.

1.An electric shock

In order to avoiding the electric shock, please wear the rubber-insulated gloves before using the tester.

2.The tester is in the testing state

When the tester is running in the testing state, the P-wire, the tested device, the test probe and the output port all carry the high voltage. Please don‟t touch!

3.Changing the tested device

After finishing testing, when you change another tested device, please check to be sure:

①The tester is in the reset state.

②The test light is not flicking.

 The figure of the voltage in the liquid-crystal display is not changing.

Especially notice: Don’t touch the high voltage probe when changing the tested device.

4.Switch on or off the power switch

Once the power is cut off, please wait for a minute if you want to start it once more.

Caution: It is very dangerous in the high voltage output state to change the power state, “on” or “off”, continuously. It also do harm to the tester.

Don’t join the output port of the high voltage with any things when switching on or off the power for it maybe cause the abnormal high voltage output which is very hazard.

5.Emergency treatment

In order to avoiding the more loss, please do according to the following way under any emergent conditions such as an electric shock, the burning of the tested device or the host:

①First cut off the power.

②Then unplug the plug of the power.

6.Trouble occurring

The following situations are very dangerous. Be careful please! Because it may still output the high voltage even you press the “RESET”.

①The test light is still going on when presses the “RESET” down.

②The voltmeter shows no readout while the test light is still going on.

When one of the situations occurs, please cut off the power and unplug its plug, and don‟t use it any more. The trouble is very dangerous. Please bring it to our company or the business-acted place to repair.

7.Trouble of the test light

The voltmeter shows the readout while the test light doesn‟t light after the “TEST” is pressed. This state may be the trouble of the test light. Please make the tester off and bring it to our company or the business-acted place to repair.


II.Technical parameter

1.Model List

MODEL


CS2670A

ACW: 5kV, 20mA

CS2670A-1

ACW: 5kV, 20mA

CS2670AX

ACW: 5kV, 20mA

CS2670Y

ACW: 5kV, 20mA

CS2671AX

ACW: 10kV, 20mA; DCW: 10kV, 10mA

CS2671BX

ACW: 10kV, 50mA; DCW: 10kV, 20mA

CS2671TX

ACW: 10kV, 100mA

CS2672BX

ACW: 5kV, 100mA

CS2672CX

ACW: 5kV, 100mA DCW: 5kV, 20mA

CS2672DX

ACW: 5kV, 20mA DCW: 5kV, 10mA

CS2673X

ACW: 5kV, 200mA

CS2674AX

ACW: 20kV, 10mA; DCW: 20kV, 10mA

2.Technical parameters

①CS2670A/CS2670A-1/CS2670AX

MODEL

CS2670A

CS2670A-1

CS2670AX

CS2670Y

 

 

 

 

 

ACW

Output voltage

0.00~5.00kV

Max. output power

100VA5.000kV/20mA

Max. current

20mA

Current range

2mA20mA

Output

waveform

sine wave

Waveform

distortion

≤5%unload or pure resistance load

Test time

0.0s999s   0=continue

Volt meter

Range

0.00kV5.00kV

Accuracy

±5%+5 word

resolution

10V

Displayed value

RMS

 

Ammeter

Range

AC

0.100mA   20.00mA

resolution

AC

2mA1uA 20mA10uA

Accuracy

±5%+5 words

 

Timer

Range

0.0s999s

Min. resolution

0.1s

Accuracy

±1%+50ms

PLC interface

no

no

yes

no

Remote interface

yes

②CS2671AX/CS2671BX/CS2671TX

MODEL

CS2671AX

CS2671BX

CS2671TX

 

 

 

 

 

 

 

 

AC W

Output voltage

0.00~10.00kV

Max.output

power

100VA

500VA

1000VA

Max. current

20mA

50mA

100mA

Current range

2mA20mA

2mA20mA50mA

2mA20mA100mA

Output

waveform

sine wave

Waveform

distortion

≤5%

Test time

0.0s999s 0.0= continue

 

 

D CW

Output voltage

0.00~10.00kV

Max. output

power

100VA

500VA

--------------------

Max. current

10mA

20mA

--------------------

Current range

2mA10mA

2mA20mA

--------------------

Ripple factor

≤5%

Test time

0.0s999s 0= continue

 

 

Volt meter

Range

0.00~10.00kV

Accuracy

±5%+5 words

resolution

10V

Displayed value

RMS

 

Ammeter

Measure Range

AC

0.100mA 20.00mA

0.100mA 50.00mA

0.100 100.0mA

DC

0.100mA 10.00mA

0.100mA 20.00mA

--------------------

Resolution

AC

2mA1uA20mA10uA,50mA100mA):0.1mA

Accuracy

±5%+5 words

 

Timer

Range

0.0s999s

Min. resolution

0.1s

Accuracy

±1%+50ms

PLC interface

no

no

Yes

Remote interface

yes

③CS2672BX/CS2672CX/CS2672DX

MODEL

CS2672BX

CS2672CX

CS2672DX

 

 

 

 

A C W

Output

voltage

0.00~5.00kV

Maxoutput

power

500VA

100VA

Max. current

100mA

20mA

Current range

2mA20mA100mA

2mA20mA

Output

waveform

sine wave

Waveform

distortion

≤5%

Test time

0.0s999s 0.0= continue

 

 

D C W

Output

voltage

-----------------------------

0.00~5.00kV

0.00~5.00kV

Max. output

power

-----------------------------

100VA

50VA

Max. current

-----------------------------

20mA

10mA

Current range

-----------------------------

2mA20mA

2mA10mA

Ripple factor

-----------------------------

≤5%

Test time

-----------------------------

0.0s999s 0= continue

 

Volt meter

Range

0.00~50.00kV

Accuracy

±5%+5 words

resolution

10V

Displayed

value

RMS

 

 

Ammeter

Measure Range

AC

0.100mA 100.0mA

0.100mA 100.0mA

0.100mA20.00mA

DC

----------------------------

0.100mA 20.00mA

0.100mA10.00mA

Resolution

2mA1uA 20mA10uA, 50mA0.1mA

Accuracy

±5%+5 words

 

Timer

Range

0.0s999s

Min. resolution

0.1s

Accuracy

±1%+50ms

PLC interface

no

no

Yes

Remote interface

yes

④CS2673X/CS2674AX

MODEL

CS2673X

CS2674AX

 

 

 

A

C

W

Output voltage

0.00~5.00kV

0.00~20.00kV

Max. output

power

500VA

200VA

Max. current

200mA

20mA

Current range

2mA20mA200mA

2mA10mA

Output

waveform

sine wave

Waveform

distortion

≤5%

Test time

0.0s999s 0.0= continue

 

 

D C W

Output voltage

--------------------------------

0.00~20.00kV

Max. output

power

--------------------------------

200VA

Max. current

--------------------------------

10mA

Current range

--------------------------------

2mA10mA

Ripple factor

≤5%

Test time

0.0s999s 0= continue

 

Volt meter

Range

0.00~5.00kV

0.00~20.00kV

Accuracy

±5%+5 words

resolution

10V

Displayed value

RMS

 

Ammeter

Measure Range

AC

0.100mA 200.00mA

0.100mA 10.00mA

DC

-----------------------------------

0.100mA 10.00mA

Resolution

2mA1uA 20mA10uA, 200mA0.1mA

Accuracy

±5%+5 words

 

Timer

Range

0.0s999s

Min. resolution

0.1s

Accuracy

±1%+50ms

PLC interface

no

no

Yes

Remote interface

yes


  • <<Previous page
  • Next page >>