I. General Information
HZ1902is a double-beam UV/VIS Spectrophotometer with a scanning function. It has a wider range of spectrum and is of high quality. Due to its built-in micro-computer system, plus excellent optical and electric circuit system, rational mechanical structure and a large LCD display screen, the instrument will surely be an effective and visual means useful to lab analysis in all kinds of industries.
HZ1902UV/VIS Spectrophotometer adopts a man-machine dialogue operation method. It is easy to learn. Menu at the large LCD screen allows every item to be chosen and accepted corresponding to each operation step. In this way, you can perform function you needed. The change of lamp is simple. By turning several screws, you change the lamp, which will be at the optimal position without being complicated light-adjustment.
Instrument is designed to be protected by software program to restore the instrument to normal work status in a failure.
Being an excellent practical UV/VIS spectrophotometer, HZ1902provides fast and simple analytical methods to be widely used in organic, inorganic, petroleum, pharmacy, environment, bio-chemistry, medicine, food and other national economy departments for QC and QA.
1.2.1 Technical Indicators
Model | HZ1902 | HZ1901 |
Photometric | Transmittance,Absorbance,Energy,Reflectivity | |
Wavelength Rang | 190nm~1100nm | |
Spectrum Bandwidth | 1nm | 2nm |
Stray Light | ≤0.03%T (220nm NaI、360nm NaNO2) | |
Wavelength Accuracy | ±0.3nm | |
Wavelength Repeatability | 0.1nm | |
Photometric Range | -4A~4A | |
Photometric Accuracy | ±0.3%T(0~100%T) ±0.002A(0~0.5A) ±0.004A(0.5~1A) | |
Photometric Repeatability | 0.001A(0~0.5A) 0.002A(0.5~1A) | |
Stability | ≤±0.0004A/h(2 hours after turn-on, at 500nm) | |
Noise | ≤0.0003A | |
Baseline Flatness | ≤±0.0008A | ≤±0.001A |
Power | AC 100V~240V 50Hz/60Hz | |
Light source | Deuterium & Tungsten-halogen lamp | |
Scanning Speed | Fast Middle Slow |
1.2.1 Use Conditions
Ambient Temperature : | 5℃~40℃ |
Ambient Humidity: | ≤80% |
Elevation | ≤2000m |
Operation Voltage: | AC 100V~240V 50Hz/60Hz |
Rated Power | 200W |
Fluctuation range of Voltage | AC 85V~265V 50Hz/60Hz |
Fuse Model/Fuse Specification | RF1-20 /F3.15AL250V |
There shall be no strong electric-magnetic interference and vibration in room. |
1.2.3 Specifications
Dimensions: | 650mm (long)×450mm (wide)×220mm (high) |
Weight: | 25kg |
There are two main functions: automatic control and analytical test of information processing.
1.3.1 Automatic Control
1. Switch on the instrument, the instrument performs self-diagnosis its internal system work status and automatic wavelength calibration.
2. Wavelength change and automatic location
3. Automatic switching of filter
4. Automatic switching of lamp
5. Automatic choice of optimal switching point of lamp souce
6. Display and print of altas and data
7. Display of error information
1.3.2 Analytical Test and Information Processing
1. Spectrum scan
2. Single wavelength quantitative analysis (including multi-valences regression concentration measurement)
3. Two and three wavelength quantitative analysis (including multi-valences regression concentration measurement)
4. Multi-valences derivative measurement
5. Multi-wavelength measurement
6. Chemical-dynamic measurement
7. Combination and operation of curves, multi-valences differential smooth of curves, reduction and amplification of alta and storage and call-in of curves.
8. Identification, searching and print of peak values