I.Overview
GA-2009 type air generator is used as a low noise compressor air source; the apparatus is provided with a two stage stabilized and automatic temperature control system, purification system by the filter. The activated carbon chamber and silica gel chamber composed of three levels of purification, can effectively remove impurities and water molecules in the air. Can be used for chromatography, flame photometer of various models at home and abroad, can replace the cylinder to make your work more convenient.
II.The Main Technical Parameters
1, working conditions: power supply voltage: 220V 50Hz
Ambient temperature: 15-40C degrees relative humidity is less than 70%
Without a lot of dust and corrosive gas pollution.
2, air purity: three grade purification without oil
3, the output flow rate: GA-2009 type 0-2L/min
4, the output pressure: A output 0-0.4Mpa The output of B 0-0.5Mpa
5, noise: less than 42 DB
6, rated power: 150W GA-2009; 170W PGA-5L
7, size: 465 x 235 x 360mm (L * W * H);
8, weight: about GA-2009 22Kg;