· To meet standard ISO7072;
· Automatic chromaticity compensation;
· Automatic range switch, several measurement units can be selected;
· 2000 sets of data can be stored;
· Auto power-off;
· Under-voltage prompt;
· Communication function through USB socket;
· Power-off protection function. When the meter is shut down, the stored calibration data and setting parameters will not be lost.
Display Unit | NTU | FNU | EBC |
Measuring Range (NTU) | (0.00 ~ 20.00), (20.0 ~ 200.0), (200 ~ 1000) Automatic Range Switch | (0.00 ~ 20.00), (20.0 ~ 200.0), (200 ~ 1000) Automatic Range Switch | (0.00 ~ 4.90), (4.9 ~ 49.0), (49 ~ 245) Automatic Range Switch |
· Zero drift: ≤±0.5%FS/30 min
· stability: ≤±0.5%FS/30 min
· Repeatability: ≤0.5%
· Indication error: ≤ ±6%
· Working Environment
a. Ambience temperature: 5℃~35℃
b. Relative humidity: ≤85%
c. Power supply: DC voltage (4.8~6)V, the supply current not less than 300mA;
d. No corrosive gas exist in the surrounding air;
e. No vibration that could affect performance exist;
f. No strong electromagnetic interference to surrounding except the geomagnetic field
· Dimensions (mm): 220×100×80 (L×W×H).
· Weight (kg): About 0.8kg.